Sturdy Structure & Superior Measurement Accuracy Suitable For OR.

Wide Screen options to choose from. Fanless design making it idle choice for silent environment or operation room.

Clear

Frequently Purchased Together

Five-Stand Trolley

Printing Paper 50mm for Northern Meditec Patient Monitors

This item is already in one of your wishlists
Add to Wishlist
SKU: NGE Categories: ,
Northern
Product Description

Hospital Patient Monitor – Gemini

Hospital patient monitor features:

• 15.6” colour TFT LCD Touch Screen
• Suitable for adult, pediatric & neonatal patients
• Flexible parameters configuration for different clinical environments
• Screen Upgradeable To 17″ & 19″ Screen Post Purchase
• Standard Configuration 3 Lead ECG, HR, RESP, NIBP, Spo2, 2 Temp, PR
• Accurate Alarming System
• Up To 10 Waveform Display
• Pacemaker Detection
• ST & Arrhythmia Analysis
• Multi Selectable Screen Options
• 8 Hours holographic waveform review
• 168 hour graphical & tabular trend review with customisable size & colour settings
• Rechargeable high capacity Li-ion battery

Optional
• Thermal Printer, Trolley, Wall Mount, Masimo/Nellcor SpO2, Dual IBP, Mainstream/Sidestream EtCO2, Multigas, BIS, NMT, 12 Lead ECG, Cardiac Output, Wifi, 3G/4G Wireless Connection, Central Monitoring System

2 Year Warranty

Kit Contents Specifications
Display
15.6"color TFT screen
Resolution: 1366.768 dpi
Number of traces: 12 waveforms

Indicator
Power indicator light
Battery indicator light

Battery
Rechargeable lithium battery

I/0
LAN: 1 standard RJ45 Port
WLAN: IEEE802.11 b/g/n
USB: 2 USB ccnnectors
SD: 1 SD card Socket
VGA: 1 standard cclor VGA monitor ccnnector
Output: 1 standard ccnnector for Nurse Call, Defib Sync Analog Output

Recorder (Optional)
Built-in, thermal dot array
Waveform: 3 channels
Paper width: 50mm
Print speed: 12.5 mm/s, 25mm/s, 50mm/s

EGG
Lead Type: 3-lead, 5-lead,12-lead
EGG waveform: 2 channels, 7 channels, 12 channels
Gain selection: x0.25, x0.5; x1; x2
Sweep speed: 6.25mm/s, 12.5mm/s, 25mrnls, 50mm/s
Filter: Diagnostic mode: 0.05-150Hz
Monitoring mode: 0.5-40 Hz
Surgery mode: 1-20Hz
Notch: 50Hz, 60Hz
Heart Rate range: Adult: 15-300bpm;
Pediatric/ Neonate: 15-350bpm;
Accuracy: ±1 bpm or ±1

Respiration
Method: Trans-thoracic impedance
Operation modes: Auto/ Manual
Measurement range: Adult: 0-120 bpm;
Neonate/Pediatric: 0-150 bpm
Apnea alarm delay: 1 Os, 15s, 20s, 25s, 30s, 35s, 40s
Apnea Alarm: Selectable
Sweep Speed: 6.25mm/s, 12.5mm/s, 25mm/s

Sp02
Measurement range: 0-100%
Resolution: 1%
Accuracy: ±2% (70-100% AduiVPediatric);
±3% (70-100% Neonate);
0-69% unspecified
Refreshing rate: 1 s

Masimo Rainbow®
SP02(0ptional)
Accuracy: ±2% (70-100% AduiVPediatric, non-motion,low perfusion);
±3% (70-100% Neonate, non-motion);
±3% (70-100% motion);
0-69% unspecified
Other measurements: SpHb®, SpOC'", Spco®, SpMet®

Pulse Rate
Range: 0-254 bpm
Resolution: 1 bpm
Accuracy: ±3bpm (non-motion)
±5bpm (motion)
Refreshing rate: 1 s

NIBP
Method: Automatic oscillometric
Operation modes: Manual/Auto/STAT
Measurement unit: mmHg/kPa selectable
Measurement types: Systolic, Diastolic, Mean
Measurement range:
Range of systolic pressure:
Adult Mode - 40 -270 mmHg
Pediatric Mode - 40 -200 mmHg
Neonatal Mode - 40 -135 mmHg
Range of diastolic pressure: Adult Mode - 10 -210 mmHg
Pediatric Mode - 10 -150 mmHg
Neonatal Mode - 10-100 mmHg
Range of mean pressure:
Adult Mode - 20 - 230mmHg
Pediatric Mode - 20 - 165mmHg
Neonatal Mode - 20 -110mmHg
Accuracy: Max mean error: ±5 mmHg
Measurement Interval: 1 ,2,3,4,5, 10, 15,30,60,90, 120,180,240,480minutes
Over-pressure protection: Double safety protection
Cuff pressure range: 0-280 mmHg
Resolution: 1 mmHg

Temperature
Accuracy: ±0.1 'C or ±0.2 •F(without probe)
Measurement range: 0-50'C (32 to 122 •F)
Channel: Two-channels
Resolution: 0.1 'C
Parameters: T1 , T2 and TD

IBP (Optional)
Channel: 2-channel I 4-channel
Measurement types: Systolic, Diastolic, Mean
Range:OmmHg-300mmHg
Resolution: 1 mmHg
Accuracy: ±2% or ±1mmHg, whichever is greater (without sensor)
Sensitivity: 5uV/mmHgN
Impedance range: 300 to 3000

C.O. (Optional)
Method: Thermodilution
Range: C.O .. 0.1 to 20Umin
TB: 23 to 43'C
T1: Oto27'C
Accuracy: C.O.:±S% or ±0.1 Urn in, whichever is greater
TB, T1 : ±0.1'C (without sensor)

Phasein ISA SidestreamC02 (Optional)
Warm-up time: Full accuracy within 10 seconds
Sampling flow rate: 50ml/min (+/-10ml/min)
Accuracy: ± (0.2 % + 2 % of the reading)
Measurement Range: 0 -15%
Rise time: 200ms, typical at 50ml/min flow rate
Total response time: within 3 seccnds (with 2 m Nomoline
sampling line)
AWRR Range: 0-150bpm
AWRR Accuracy: ±1 breath

Phasein IRMA Mainstream C02 (Optional)
Measurement Range: 0 -15%
Warm-up time: Full accuracy within 10 seconds
Accuracy: ± (0.2 % + 2 % of the reading)
AWRR Range: 0-150bpm
AWRR Accuracy: ±1 breath

Phasein Multi-gas IRMA AX+Mainstream (Optional)
Gas: C02, N20,HAL, ISO, ENF, SEV, DES with automatic identification
Warm-up time: Full accuracy within 20 seconds for IRMA AX+
C02 Accuracy . 0-10%: ± (0.2 % + 2 % of the reading)
10-15%: ± (0.3% + 2% of the reading)
N20Accuracy: 0-100%,± (2% + 2% of the reading)
HAL, ISO, ENF: 0-8% ' ± (0.15 % +5 % of the reading)
SEV: 0-1 O"k ± (0.15 % +5 % of the reading)
DES:0-22%: ± (0.15 % +5 % of the reading)
Agent identification time:< 20s (typical < 10s)
AWRR range: 0-150bpm
AWRR accuracy: +/-1 bpm
Apnea time: 20s, 25s, 30s, 35s,40s

Operation Environment
Power AC 1 00-250V, 50/60Hz
Temperature 0-40'C
Humidity 15-85%
Patient Range Adult, Pediatric, Neonate