Display
- 12.1″ color TFT LCD touch Screen,resolution: 800 x 600
- 8 waveform display, 12-leads ECG display
I/O
- LAN: 1 standard RJ45 port
- WLAN: IEEE802.11 b/g/n
- USB:2 USB connectors
- SD: 1 SO card socket
- Output:1 connector for nurse call, Defib sync analog output
- VGA: 1 VGA external socket (optional)
ECG
- Lead type :3-lead,S-Iead, 12-lead
- ECG waveform:2 channels,7 channels, 12 channels
- Display sensitivity:
- 2.5mm/mV (x0.25), 5mm/mV (x0.5), 1 Omm/mV (x 1.0), 20mm/mV (x2.0)
- Wave sweep speed: 6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
- Bandwidth
- Diagnostic mode: 0.05Hz-100Hz
- Monitor mode: 0.5Hz-40Hz
- Surgery mode: 1Hz-20Hz
- Strong filter mode: 5Hz-20Hz
- CMRR>100dB
- Notch: 50/60Hz notch filter can be set to on or off
- Differential input impedance> SMO
- Electrode polarization voltage range: ±400mV
- Baseline recovery time<3s after defibrillation (in monitor and surgery mode)
- Calibration signa1:1 mV (peak – peak), accuracy ±3%
TEMP
- Accuracy: ±0.1 C or ±0.2 C F (without probe )
- Measurement range: 5-50’C( 41- 122′ F )
- Channel : Two channels
- Resolution: 0.1 C
- Parameters : T1,T2 and TO
RESP
- Method:Trans-thoracic impedance
- Operation modes: Auto/Manual
- Measurement range: Adult:0- 120 bmp;
- Neonate/Pediatric:0-150bpm
- Apnea alarm delay:1 Os, 1Ss.20s,25s,30s,35s,40s
- Apnea alarm: Selectable
- Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s
NIBP
- Measurement method : Automatic oscillometric method
- Operating mode:Manual, automatic, continuous
- Measurement unit: mmHg/kPa selectable
- Typical measurement time: 20-40s
- Measurement type: Systolic, Diastolic, Mean
- Measurement range {mmHg)
Range of Systolic pressure:
- Adult 40-270
- Pediatric 40-230
- Neonatal 40-135
Range of Diastolic pressure:
- Adult 10-210
- Pediatric 10-150
- Neonatal 10-95
Range of Mean pressure:
- Measurement accuracy
- Adult 20-230
- Pediatric 20-165
- Neonatal 20-105
Measurement accuracy
- Maximum average error: ±5mmHg
- Maximum standard deviation: 8mmHg
- Resolution: 1 mmHg
- lnterval:1,2,3,4,5, 10, 15,30,60,90, 120, 180,240,480 minutes
- Overpressure protection: Software and hardware, double safety protection
- Cuff pressure range: 0-280mmHg
Northern SpO2
- Measurement range : 0·1 00%
- Resolution: 1%
- Accuracy: ±2% ( 70-100%, Adult/Ped iatric);
- ±3% { 70-100% , Neonate);
- 0 -69%,unspecified
- Refreshing Rate: 1 s
Masimo SpO2
- Measurement range : 0-100%
- Resolution: 1%
Accuracy:
- ±2% { 70-100% , Adult/Pediatric,non-motion,low prelusion);
- ±3% { 70-100%, Neonate,non-motion);
- ±3% ( 70-100% , motion);
- 0-69%,unspecified
- Refreshing Rate: 1s
Pulse Rate
- Range: 30-254 bpm
- Resolution: 1 bpm
- Accuracy: ±2bpm (non-motion) ±5bpm (motion)
- Refreshing rate: 1s
IBP (Optional)
- Channel:2-channel or 4-channel
- ART: 0 to 300 mmHg
- PA: -6 to 120 mmHg
- CVP/RAP/LAP/ICP: -10 to 40 mmHg
- Measurement range: P1/P2 -50 to 300 mmHg
- Resolution:1 mmHg
- Accuracy: ±2% or ± 1 mmHg, whichever is greater (without sensor)
- Sensitivity: 5uV /mmHg/V
- Impedance range: 300 to 3000
Phasein IRMA Sidestream CO2 (optional)
- Warm-up time: Full accuracy within 10 seconds
- Sampling flow rate: 50ml/min( +/-1 0/min)
- Accuracy: 0-15% (±0.2% of the reading) 15-25%, unspecified
- Measurement Range: 0 -25%
- Rise time: 200ms,typical at 50ml/min flow rate
- Total response time: within 3 seconds(with 2m Momoline sampling line)
- AWRR Range: 0-150bpm
Phasein IRMA Mainstream CO2
- Measurement Range: 0 -25%
- Warm-up time: Full accuracy within 10 seconds
- Accuracy: 0-15% (±0.2% of the reading)
- 15-25%, unspecified
- AWRR Range: 0-150bpm
C.O. (optional)
- Method: Thermodilution
- Range: C.O.: 0.2 to 20 L/min
- TB: 23 to 45 ·c
- T1 : -1 to 27C
- Accuracy:C.O.:±5% or ±0.1 L/min, whichever is greater
- TB, T1 : ±O5C (without sensor)
Recorder (optional)
- Built-in, Thermal dot array
- Horizontal resolution :16 dots/mm (25 mm/s paper speed)
- Vertical resolution:8 dots/mm
- Paper speed:25 mm/s, 50 mm/s
- Number of waveform channels: 3
Operating Environment
- Power: AC-=250, 50/60Hz
- Temperature:5-40C
- Humidity: <80%
- Patient Range: Adult, Pediatric, Neonate
There are no reviews yet.