The Northern Aquarius plus is our most popular patient monitor bringing all major parameters to a convenient little monitor. Includes NIBP, SpO2, ECG, Surface Temp, PR, EtCo2Β and Touchscreen. Suitable for Adult, Pediatric and Neonate patients.
Featuring a suite of essential parameters and a user-friendly 8β colour TFT LCD touchscreen, the Northern Aquarius Plus is a reliable, economic choice for healthcare facilities seeking accuracy, portability, and ease of use.
Key Features:
-
Broad Patient Compatibility: Suitable for adult, paediatric, and neonatal patients, ensuring versatile use across different healthcare scenarios.
-
Comprehensive Monitoring: Standard configuration includes 3 Lead ECG, NIBP, SpO2, and Temp, providing critical vitals at a glance.
-
High-Quality Display: 8β colour TFT LCD screen with touchscreen capability for intuitive navigation and data entry.
-
Extended Battery Life: Equipped with a Li-ion battery offering up to 10 hours of operation, ideal for continuous monitoring.
-
Portable Design: Ergonomic, lightweight, and sturdy, making it easy to move and position where needed.
-
Advanced Data Analysis: Spot-check and continuous monitoring modes with alarms, plus graphical and tabular trend review for detailed patient data analysis.
-
Customisable Parameters: Flexible configuration for different clinical environments, with optional features including Thermal printer, Wall Mount, EtCO2, Multigas, and more for enhanced functionality.
Β
Display
8β color TFT LCD Screen,resolution: 800 x 600
ECG
Lead type
3-lead:l, II, Ill
SΒ· lead:l, II, lll,aVR, a VI., aVF, V
Display sensitivity:
2.5mm;mv (x0.25), Smm;mv (x0.5), 10mm;mv (x1.0),
20mm;mv {x2.0)
Wave sweep speed: 6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s
Bandwidth
Diagnostic mode: 0.05Hz-100Hz
Monitor mode: 0.5Hz-40Hz
Surgery mode: 1Hz-20Hz
Strong filter mode: 5Hz-20Hz
CMRR>100dB
Notch: 50/60Hz notch filter can be set to on or off
Differential input impedance> SMO
Electrode polarization voltage range: Β±400mV
Baseline recovery time<3s after defibrillation {in monitor and
surgery mode)
Calibration signal:1mV (peak β peak), accuracy Β±3%
RESP
Measurement method : Thoracic electrical bio1 pedance
Measuring lead: Lead I, II
Wave gain: x0.25, xO.S, x 1, x2
Respiratory impedance range: 0.5 Β·50
Baseline impedance: 500-40000
Gain: 10 grades
Scan speed: 6.25mm/s, 12.5 mm/s, 25mm/s
TEMP
Measurement method: Thermistor
Measuring range: 5- SOβC (41-122βF)
Resolution: 0.1βC
Measurement accuracy: Β±0.1βC
Printing (optional)
Built-in,Thermal dot array
Horizontal resolution :16 dots/mm (25 mm/s paper speed)
Vertical resolution:8 dots/mm
Paper speed:25 mm/s, 50 mm/s
Number of waveform channels:3
NIBP
Measurement method : Automatic oscillometric method
Operating mode:Manual, automatic, continuous
Measurement unit: mmHg/kPa selectable
Typical measurement time: 20-40s
Measurement type: Systolic, Diastolic,Mean
Measurement range {mmHg)
Range of Systolic pressure:
Adult 40-270
Pediatric 40-230
Neonatal 40-135
Range of Diastolic pressure:
Adult 10-210
Pediatric 10-150
Neonatal 10-100
Range of Mean pressure:
Measurement accuracy
Adult 20-230
Pediatric 20-165
Neonatal 20-110
Maximum average error: Β±SmmHg
Maximum standard deviation: 8mmHg
Resolution: 1 mmHg
interval:1 ,2,3,4,5, 10, 15,30,60,90, 120, 180,240,480minutes
Overpressure protection: Software and hardware,double
safety protection
Cuff pressure range: 0-280mmHg
Northern SpO2
Measurement range : 0Β·1 00%
Resolution: 1%
Accuracy: Β±2% ( 70-100%, Adult/Pediatric);
Β±3% { 70-100% , Neonate);
0 -69%,unspecified
Refreshing Rate: 1 s
Masimo SpO2 (optional)
Measurement range : 0-100%
Resolution: 1%
Accuracy: Β±2% { 70-100% , Adult/Pediatric,non-motion,low prelusion);
Β±3% { 70-100%, Neonate,non-motion);
Β±3% ( 70-100% , motion);
0-69%,unspecified
Refreshing Rate: 1
Infrared Ear thermometer (optional)
Diplayed range: 34- 42.2βC {93.2-108 Fβ)
Operation ambient temperature range: 10-40βC{ 50-104βF)
Accuracy for displayed temperature range:
~35βC{ 95.9βF) -s42.2βC{ 107.6βF) range Β±0.2βC{ 0.4βF)
< 35βC ( 95.9βF ) β ~34βC { 93.2βF) range Β±0.3βC{ O.SβF)
Phasein IRMA Sidestream CO2 (optional)
WarmΒ·up time: Full accuracy within 10 seconds
Sampling flow rate: SOml/min(+/Β·10/min)
Accuracy:Β±(0.2%+2% of the reading)
Measurement Range: 0 Β·15%
Rise time: 200ms,typical at SOml/min flow rate
Total response time:
within 3 seconds(with 2m Momoline sampling line)
AWRR Range: 0Β·150bpm
AWRR Accuracy:Β±1 breath
Phasein IRMA Mainstream CO2 (optional)
Measurement Range: 0 Β·15%
Warm-up time: Full accuracy within 10 seconds
Accuracy:Β±(0.2%+2% of the reading}
AWRR Range: 0-150bpm
AWRR AccuracyΒ·+ 1 breath
Operating Environment
Phasein IRMA MAinstream CO2
Power: AC-=250, 50/60Hz
Temperature: 0-40C
Humidity: 15-85%
Patient Range: Adult, Pediatric, Neonate