Northern Venus Operating Patient Monitor
Northern Venus Operating Patient Monitor may not be in stock. Please call us for ETA before purchasing.
Couldn't load pickup availability
The Northern Venus Operating Patient Monitor is a premier patient monitoring system designed to support the dynamic requirements of operating rooms. Featuring a large 15.6” colour TFT LCD touchscreen, it provides exceptional clarity and ease of use for monitoring adult, paediatric, and neonatal patients. Its flexibility in parameter configuration allows it to be seamlessly integrated into various clinical environments, making it an indispensable tool for surgical teams.
Key Features:
- Broad Patient Compatibility: Designed for adult, paediatric, and neonatal monitoring, offering versatility in patient care.
- Enhanced Visual Display: A 15.6” colour TFT LCD touchscreen offers superior visibility and easy navigation through patient data.
- Comprehensive Monitoring Suite: Standard configuration includes 3 Lead ECG, HR, RESP, NIBP, SpO2, 2 Temp, and PR, equipped with pacemaker detection and ST & Arrhythmia analysis.
- Advanced Visualisation Options: Supports up to 10 waveform displays with multiple screen options for in-depth monitoring.
- Extended Review Capabilities: Features 8 hours of holographic waveform and 168 hours of graphical & tabular trend review with adjustable settings.
- Reliable Operation: Comes with a rechargeable high-capacity Li-ion battery for uninterrupted monitoring.
- Expandable Functionality: Offers optional upgrades like thermal printer, trolley, wall mount, and advanced monitoring options for comprehensive patient assessment.
- Seamless Connectivity: Includes options for WiFi, 3G/4G wireless connections, and central monitoring system integration.
Physical Specifications |
Display 15.6" TFT Touch screen Resolution: 1366 x 768 Number of traces: 10, up to 12 ECG waveforms Dimension: 398×302×183mm(WxH×D) Weight: < 7 kg under standard configuration LAN: 1 standard RJ45 port WLAN:IEEE 802.11b/g/n USB: 2 USB connectors HDMI: 1 HDMI monitor connnector Output:1 connector for Nurse call,Defib Sync Analog Output |
ECG |
Lead type :3-lead,5-lead,12-lead(optional) ECG waveform:2 channels,7 channels, 12 channels Display sensitivity(wave gain): 1.25mm/mV(×0.125), 2.5mm/mV (×0.25), 5mm/mV (×0.5), 10mm/mV(×1.0), 20mm/mV (×2.0), 40mm/mV (×4.0), Auto Wave sweep speed: 3.125mm/s, 6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s Bandwidth Diagnostic mode: 0.05Hz~150Hz Monitor mode: 0.5Hz~40Hz Surgery mode: 0.5Hz~25Hz Strong filter mode: 5Hz~25Hz CMRR>100dB Notch: 50/60Hz notch filter can be set to on or off Differential input impedance>5MΩ Electrode polarization voltage range: ±400mV HR range: 15 - 350 bpm Baseline recovery time <3s after defibrillation (in monitor and surgery mode) Calibration signal:1mV (peak - peak), accuracy ±3% |
RESP |
Measurement method : Thoracic electrical bioimpedance Measuring lead: Lead I, II Wave gain: x0.25, x0.5, x1, x2 Respiratory impedance range: 0.5-50 Respiration range: 0 - 150bpm Baseline impedance: 500-40000 Gain: 10 grades Scan speed: 3.125mm/s, 6.25mm/s, 12.5 mm/s, 25mm/s |
TEMP |
Accuracy:±0.1C or ±0.2°F (without probe) Measurement range: 5~50C (41~122°F) Channel: Two channels Resolution: 0.1C |
SpO2 |
Measurement range : 0-100% Parameter monitoring: Perfusion Index(PI), Pleth Variability Index(PVI) Resolution: 1% Accuracy: ±2% or ±2bpm Refreshing Rate: 1s Pleth wave speed: 3.125mm/s, 6.25mm/s, 12.5 mm/s, 25mm/s |
Masimo SET SpO2 (Optional) |
Measurement range : 0-100% Resolution: 1% Accuracy: ±2% (70-100%, Adult/Pediatric,non-motion, low perfusion); ±3% (70-100%, Neonate, non-motion); ±3% (70-100%, motion); 0-69%,unspecified Refreshing Rate: 1s |
Pulse Rate |
Range: 30~300 bpm Resolution: 1bpm Accuracy: ±2bpm (non-motion) ±5bpm (motion) Refreshing rate: 1s |
NIBP |
Measurement method: Automatic oscillometric method Operating mode: Manual, automatic, continuous Measurement unit: mmHg/kPa selectable Typical measurement time: 20~40s Measurement type: Systolic, Diastolic, Mean Measurement range: (mmHg) Range of Systolic pressure: Adult 40-280, Pediatric 40-200, Neonatal 40-135 Range of Diastolic pressure: Adult 10-210, Pediatric 10-150, Neonatal 10-95 Range of Mean pressure: Adult 20-230, Pediatric 20-165, Neonatal 20-105 Measurement accuracy Maximum average error: ±5mmHg Maximum standard deviation: 8mmHg Resolution: 1mmHg Interval:1,2,3,4,5,10,15,30,60,90,120,180,240,480minutes Overpressure protection: Software and hardware, double safety protection Cuff pressure range: 0-300mmHg |